BUNNELL, INC. LIFE PULSE; VENTILATOR, HIGH FREQUENCY, CIRCUIT Back to Search Results
Model Number 203
Device Problems Overheating of Device (1437); Overfill (2404)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/02/2019
Event Type  malfunction  
Event Description
The heater water chamber on the jet alarmed that chamber was too full and would not stop.Called (b)(6) and spoke with representative.He advised respiratory therapist (rt) to empty some water from the chamber by removing the green tubing from the gas outlet, lifting the water cartridge out of its holder and empty some of the water from the chamber back through the green tubing.This worked for a while, but then the jet started alarming that the water chamber was too hot.At this point it appeared that there was very little water in the chamber.Rt called (b)(6) again and spoke with the same representative, who advised that he change out the jet, which we did.Then, the jet appeared to run without difficulty.No harm to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFE PULSE
Type of Device
VENTILATOR, HIGH FREQUENCY, CIRCUIT
Manufacturer (Section D)
BUNNELL, INC.
436 lawndale dr.
s salt lake UT 84115
MDR Report Key8662375
MDR Text Key146794568
Report Number8662375
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 05/21/2019,05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number203
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/21/2019
Date Report to Manufacturer06/03/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/03/2019
Type of Device Usage N
Patient Sequence Number1
-
-