Literature article entitled: "outcomes of a metal-on-metal total hip replacement system".The study analyzed the outcomes of metal on metal thr systems involving the corail stem, pinnacle cup, ultramet liner, and articuleze femoral head.578 uncemented hips were followed in the study.39 hips from 38 patients were revised within a mean time of 3.5 years.They were identified by age/gender.Patient #34 is a (b)(6) year old female revised for adverse reaction to metal debris.Elevated cocr blood metal ions were noted prior to revision.The head and liner were revised.Intraoperative findings include mild effusion, necrotic tissue in capsule, and trunnion corrosion.
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|