The patient was undergoing a thrombectomy procedure to treat a femoral popliteal lesion using an indigo system separator 6 (sep6).During the procedure, the physician advanced the sep6 through an indigo system aspiration catheter 6 (cat6).The physician then began retracting and advancing the sep6 within the cat6.However, while making a pass with the sep6, the tip of the sep6 became dislodged within the cat6.The physician decided to end the procedure and remove the sep6 and cat6 together from the patient.No additional information is available.
|
Results: the (b)(6) distal tip was bent, had offset wrapping coil winds, and the bulb was missing.Conclusions: evaluation of the returned (b)(6) confirmed the bulb was missing on the distal tip.If the device is repeatedly manipulated through tough clot burden, the distal tip may fatigue.If the fatigued device is continuously manipulated, the tip may become damaged such that the bulb detaches from the wire.During functional testing, the returned (b)(6) and a demonstration (b)(6) were advanced and retracted through the returned cat6 without issue.Penumbra separators are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|