Model Number 1884004
Device Problems Material Separation (1562); Vibration (1674)
Patient Problem Radiation Exposure, Unintended (3164)
Event Date 05/09/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A manufacturer representative present during the case reported that the surgeon was performing an endoscopic surgery to remove polyps when the blade started to shake and realized that the inner blade broke off.Another blade from the same batch was used and the exact same thing occurred.One of the inner blades was missing but intraoperative x-ray confirmation validated that it was not inside the patient.It was noted that the other tip was found.The procedure was performed with the third blade available from the same batch which worked correctly.There was a 50-minute delay in the procedure.The patient was alive with no injury.
 
Manufacturer Narrative
Analysis found that visually, the distal tips of the inner cutter broke off which would have resulted in the reported event.The breakage occurred at the first proximal tooth valley.The portion of the tip that was returned, measured approximately 0.24¿ in length (the other tip not returned would have measured approximately the same length.The outer assemblies were deformed in such a way that indicates the inner blade teeth impacted the outer teeth at the 2nd proximal valley while moving in a ccw direction which resulted in the observed breaks.It could not be determined if the expanded tip outside diameters had been machined / turned.There were no signs of bends or concentricity issues.There were no signs of misuse or mishandling however there was some gouging on the inner shafts outside diameter 0.56¿ from the distal face of the inner hubs.There was uneven wear at the cutting tip which may indicate an interference fit between the inner and outer assemblies at the tip.(b)(4) are no longer applicable.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICUT® STRAIGHT SHAFT 4MM 11CM LONG
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
christy cain
6743 southpoint drive north
jacksonville, FL 32216
9043328353
MDR Report Key8664475
MDR Text Key146858662
Report Number3004209178-2019-10759
Device Sequence Number1
Product Code EQJ
UDI-Device Identifier00681490047456
UDI-Public00681490047456
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1884004
Device Catalogue Number1884004
Device Lot NumberHG2P0GF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2019
Initial Date FDA Received06/03/2019
Supplement Dates Manufacturer Received07/05/2019
Supplement Dates FDA Received07/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight72
-
-