MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT4
Device Problem Mechanical Problem (1384)
Patient Problem Blurred Vision (2137)
Event Date 04/22/2019
Event Type  Injury  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Additional information has been requested.(b)(4).
 
Event Description
A facility representative reported that three weeks following an intraocular lens (iol) implant procedure, the iol was exchanged for another lens due to blurred vision.The clinical reason given for the explant was "mechanical complication of iol." additional information has been requested.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional information was received indicating the toric marker was not lined up where it was planned to be, and the patient ending up with residual astigmatism.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
MDR Report Key8666652
MDR Text Key146946988
Report Number9612169-2019-00132
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.200
Device Lot Number21219953
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/08/2019
Initial Date FDA Received06/04/2019
Supplement Dates Manufacturer Received06/04/2019
Supplement Dates FDA Received07/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DISCOVISC OVD; MONARCH III D CARTRIDGE; MONARCH III INJECTOR
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-