The investigation determined that there were no device malfunctions.A partial guide was created as the doctor had submitted a partial model.The partial guide registration was reviewed and determined to be acceptable per procedural guidelines.The guide passed its quality analysis with no adjustments necessary.As the guide was not returned after several requests, manufacturing analysis is limited to production records and processes which were determined to be acceptable.The treatment plan was reviewed which shows that the implant site surface appeared to be significantly less dense than the area 7 mm deeper.The area which is deeper and denser appears to form a point.The drill may have hit the edge of this point, potentially causing it to slip and veer either mesial or distal and causing the trajectory to deviate from the treatment plan.Therefore, the most likely cause of this issue is drill slippage due to challenging patient morphology.
|