Model Number V3.0
Device Problem Computer Software Problem (1112)
Patient Problem No Patient Involvement (2645)
Event Date 05/16/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
The customer reported that a software issue was observed which resulted in the machine freezing.This occurred before a procedure.There was no patient impact.There were no error messages displayed.
 
Manufacturer Narrative
The system was examined and the reported event was replicated.The lithium battery was found to be non-conforming and was subsequently replaced to resolve the issue.The system was tested and found to meet product specifications.The manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to lithium battery.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8676957
MDR Text Key147577745
Report Number2028159-2019-01038
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV3.0
Device Catalogue Number8065752160
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/15/2019
Initial Date FDA Received06/06/2019
Supplement Dates Manufacturer Received07/12/2019
Supplement Dates FDA Received08/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-