• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH36
Device Problems Melted (1385); Component Missing (2306); Failure to Cut (2587); Detachment of Device or Device Component (2907); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number and no non-conformances were identified.
 
Event Description
It was reported that during a colectomy procedure, the scissor took more time than usual to cut.No particular mechanical resistance on the trigger.Then, after 10 minutes of use, the plastic attachment melted and got detached.The white plastic of the superior jaws was completely missing.The plastic attachment fell into the patient then was removed and kept.There was no patient consequence.
 
Manufacturer Narrative
(b)(4).Batch # r94p39.Investigation summary: the analysis results found that the device was received with the tissue pad detached and returned in a plastic bag but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11, and the device did activate during functional testing.The device was disassembled to inspect the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8684832
MDR Text Key148025024
Report Number3005075853-2019-19592
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014621
UDI-Public10705036014621
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Catalogue NumberHARH36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2019
Initial Date Manufacturer Received 05/20/2019
Initial Date FDA Received06/10/2019
Supplement Dates Manufacturer Received05/20/2019
Supplement Dates FDA Received06/28/2019
Patient Sequence Number1
-
-