• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 9X137MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 9X137MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Flaked (1246); Peeled/Delaminated (1454)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/16/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial hip procedure.It was noted the surgeon broached to 9 stem but knew it was distally tight so he touched with a 10 broach.When he went to implant the stem, it sat very proud; he only tapped it 2 or 3 times.He pulled it out to broach it again and noticed the stem left coating in the femur.The surgeon thought the stem appeared over sprayed and that something was wrong.A new stem was opened and used without an issue.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was not confirmed.The returned products were visually and dimensionally inspected in accordance with documented procedures were found to be within specification.Review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 9X137MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8685164
MDR Text Key147564154
Report Number0001825034-2019-02499
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K101086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-104090
Device Lot Number6490069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2019
Initial Date FDA Received06/10/2019
Supplement Dates Manufacturer Received07/10/2019
Supplement Dates FDA Received08/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
BIOMET STEM CATALOG#51-104090 LOT#6191175; UNKNOWN CUP; UNKNOWN HEAD; UNKNOWN LINER
Patient Outcome(s) Other;
-
-