• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRAXX NEPHROSTOMY BALLOON AND SET; LJE CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRAXX NEPHROSTOMY BALLOON AND SET; LJE CATHETER, NEPHROSTOMY Back to Search Results
Catalog Number UNBS-10-15-CS
Device Problem Deflation Problem (1149)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Occupation: or pyxis system specialist.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
It is reported during an unspecified procedure using an ultraxx nephrostomy balloon and set, the balloon failed to stay inflated during the procedure.The staff pulled a second one which performed as expected.No adverse effects to the patient have been reported as a result of this alleged product malfunction.Additional details have been requested regarding the patient and the event.At this time, no additional information has been provided.
 
Manufacturer Narrative
Investigation - evaluation: reviews of the documentation, instructions for use (ifu), and quality control were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.The lot number of the device is not known; accordingly, a review of the device history record could not be conducted.Furthermore, a review of the quality control procedures was conducted, and no gaps were discovered.The device is shipped with instruction for use (ifu) which notes: ¿always inflate the balloon with a sterile liquid.Never inflate with air, carbon dioxide or any other gas.Do not overinflate.Using excessive pressure to inflate the balloon on this device can cause the balloon to rupture¿do not exceed the maximum rated burst pressure for this balloon device.Do not pre-inflate the balloon.¿ based on the information provided and no product returned, investigation has concluded that a root cause for this event could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAXX NEPHROSTOMY BALLOON AND SET
Type of Device
LJE CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8692236
MDR Text Key147781259
Report Number1820334-2019-01353
Device Sequence Number1
Product Code LJE
Combination Product (y/n)N
PMA/PMN Number
K024050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNBS-10-15-CS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/28/2019
Initial Date FDA Received06/12/2019
Supplement Dates Manufacturer Received06/20/2019
Supplement Dates FDA Received07/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-