The device was evaluated upon receiving the device.It was confirmed that the tip of the dilator had broken off of the device.This is a very small instrument with a malleable shaft; if all of the force is placed on the tip of the instrument while bending damage to the weld could occur.A sampling of each lot in inventory were reviewed for any indication that the weld holding the tip was not strong enough.All instruments had secure intact welds with no indication of a non-conformance.There has been a total of (b)(4) sold of this product since 2012 with no additional complaints recorded for this device.This can be seen as an isolated event.Based on the information provided, this can be seen as the final report.If additional information is obtained that alleges additional patient involvement or the need for any additional corrective actions, a follow-up report will be submitted.
|