During preparation for a thrombectomy procedure, the hospital staff noticed that the tip of an indigo system aspiration catheter 6 (cat6) appeared crushed before being taking out of the package.The damaged cat6 was found prior to use and, therefore, was not used in the procedure.The procedure was completed using a new cat6.
|
Results: the returned cat6 was kinked at approximately 45.0 cm, 128.0 cm, 131.0 cm and 132.0 cm from the hub.The device was ovalized at approximately 134.0 cm from the hub.Conclusions: evaluation of the returned cat6 revealed an ovalization on the distal tip.If the device is forcefully gripped or pinched during preparation for a procedure, damage such as an ovalization may occur.Further investigation revealed kinks along the length of the returned cat6.These kinks were likely incidental to the complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|