RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2019.Date of report: 18june2019.A follow-up report will be submitted once the investigation has been complete.
|
|
Event Description
|
Customer contacted technical support (ts) stating that unit had error code messages of 3.3v supply failed, 5v supply failed, and 35v supply failed.Manufacturer's product support engineer (pse) also found fan heating.
|
|
Manufacturer Narrative
|
G4: 08oct2019; b4: 08oct2019.The device was evaluated by the field service engineer and the reported issue was confirmed.The field service engineer replaced the motor controller printed circuit board assembly (pcba) to address the issue.The issue has been resolved, the device has been tested, and the device now functions as intended.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
Customer contacted technical support (ts) stating that unit had error code messages of 3.3v supply failed, 5v supply failed, and 35v supply failed.Manufacturer's product support engineer (pse) also found fan heating.
|
|