One 7209497 arthropierce 45 degree right configured device was used for treatment and returned for evaluation.This is a four year old reusable instrument.There is symptoms of twisting and torque noted.The actuator has been sprung.The nose has been twisted and fractured.This applies during use, washing and sterilizing.The instruments are subject to friction and entanglement with baskets, trays and other devices during sterilization loading and unloading.Inspection prior to use is required.If any damages are found, the device should be removed from use and returned promptly.There are no other complaints for this lot related to this observation.No root cause related to the manufacture of this device was determined.
|