• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/06/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch: t5aw9g.Investigation summary: the analysis showed that the ats45 device was received with no apparent damage and with no reload present on the device.In addition, it was received two tr45w reloads were received.The reloads (b & c) were received partially fired 1/10 and with the reload lockout spring damaged.The returned device was tested for functionality with a test reload and it fired, cut and formed all the staples as intended.The device fired without any difficulties, the staple line was complete, and the staples meet the staple form release criteria.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lock out.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.Reload b: tr45w, r59t5z partially fired 1/10, lockout damage.Reload c: tr45w, p5907u partially fired 1/10, lockout damage.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.
 
Event Description
It was reported that during a laparoscopic appendectomy the device was fired and the blade did not advance.There was no cutting or stapling.The device opened without any problems and the procedure was completed with an alternate like device with no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8716019
MDR Text Key148558195
Report Number3005075853-2019-19849
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036001072
UDI-Public10705036001072
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/21/2024
Device Catalogue NumberATS45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2019
Initial Date Manufacturer Received 05/27/2019
Initial Date FDA Received06/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-