Catalog Number 306499
Device Problem Leak/Splash (1354)
Patient Problem No Information (3190)
Event Date 06/05/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: five samples were received.One sample has 7.5ml of blood in it and the plunger is out of the syringe.Another sample has 1.5ml of blood and the plunger rod is out of the syringe.There are other three samples with the plunger rod assembled: two have the plunger rod- rubber stopper all the way down and no tip cap; the other has the plunger rod-rubber stopper at 5ml it has the tip cap and a solution in it.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: confirmed: bd was able to duplicate or confirm the customer's indicated failure mode with the samples provided.This is the 3rd complaint for the lot# 8348869 for the same defect or symptom.Previous complaint (b)(4).There was no documentation of issues for the complaint of batch 8348869 during the production run.Root cause description: failure mode is verified, however, the posiflush product is designed to flush the iv lines, not for drawing blood.This would be considered off label use.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the stopper separated from the plunger in the bd posiflush¿ sf saline syringe during the aspiration.The following information was provided by the initial reporter: "the plunger is coming ¿unscrewed¿ and they screw it back and it sometimes works and other times pops back off when aspirating.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SF SALINE SYRINGE
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8716475
MDR Text Key149226075
Report Number1911916-2019-00608
Device Sequence Number1
Product Code NGT
UDI-Device Identifier00382903064991
UDI-Public00382903064991
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2021
Device Catalogue Number306499
Device Lot Number8348869
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/05/2019
Initial Date FDA Received06/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-