During preparation for a thrombectomy procedure, the physician found that an indigo system cat5 aspiration catheter (cat5) was broken into two pieces upon removal from the packaging.The damage to the cat5 was found prior to use.Therefore, the cat5 was not used in the procedure.The procedure was completed using a new cat5.
|
Results: the indigo system cat5 aspiration catheter (cat5) was fractured approximately 77.0 cm from the hub.The device was ovalized approximately 127.5 cm from the hub.Conclusions: evaluation of the returned cat5 confirmed a fracture.If the device is forcefully manipulated at extreme angles during removal from its packaging hoop, damage such as a fracture may occur.Further evaluation revealed a distal ovalization.This damage was likely incidental to the reported complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|