Catalog Number CAT02438
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2019
Event Type  malfunction  
Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that the cannula cracked during the procedure.
 
Event Description
It was reported that the cannula cracked during the procedure.
 
Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: seam where polymer tip meets metal shaft of flowport cannula broke.Probable root cause: design - inadequate material selection to support movement/manipulation by user - inadequate molding/assembly design, poor strength of design manufacturing - cannula not assembled, molded or machined to specification application - excessive force the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWPORT II CANNULA WITH OBTURATOR STRYKER 165MM
Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8716919
MDR Text Key148931121
Report Number0002936485-2019-00252
Device Sequence Number1
Product Code NBH
UDI-Device Identifier07613252633181
UDI-Public07613252633181
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCAT02438
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/21/2019
Initial Date FDA Received06/20/2019
Supplement Dates Manufacturer Received05/21/2019
Supplement Dates FDA Received08/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-