MAUDE Adverse Event Report: HILL-ROM BATESVILLE REFURBISHED RITEHITE; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE REFURBISHED RITEHITE; BED, AIR FLUIDIZED Back to Search Results
Model Number PR0800010019
Device Problem Device Slipped (1584)
Patient Problem No Patient Involvement (2645)
Event Date 06/10/2019
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the brake casters needed to be replaced.Per the hillrom user manual, warning: patients may use the bed for support while entering or exiting; if the unit moves unexpectedly, patient injury could occur.When the unit is unattended, ensure that both brakes are locked.The brakes for the clinitron bed are located at the right, head end and the left, foot end of the unit.To apply the brakes, step on the lower end of the brake lever to lock the wheels.To release the brakes, apply inward pressure to the upper end of the brake lever.The technician stated the bed will not be repaired and is going to be scrapped.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from the account stating the brakes were not holding.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED RITEHITE
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key8717334
MDR Text Key149135616
Report Number1824206-2019-00248
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPR0800010019
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/10/2019
Initial Date FDA Received06/20/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-