• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CALDERA MEDICAL INC. DESARA; PELVIC MESH SLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CALDERA MEDICAL INC. DESARA; PELVIC MESH SLING Back to Search Results
Catalog Number CAL-DS01
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2018
Event Type  Injury  
Manufacturer Narrative
According to the complaint, "the person helping cut the sheath right at the exit incision when dr.(b)(6) was tensioning her sling.The sheath was not held on to and was "sucked" back into the incision.Based on the information available, the cause of the reported event described is user error.The surgical technique used by this particular surgeon and assist did not follow caldera desara instructions for use (ifu# 10-139-03 rev b - ref: warning# 6 and #13).This ifu's section "surgical implant technique" specifically instructs the surgeon "cut off the ends of the mesh assembly medial to the tip and suture connection system and pull off the sheath." additionally, the use of this device is restricted to physicians trained in performing sub-urethral sling procedures for treating stress urinary incontinence.Although not intended for long-term use, it appears that the sleeve portion of the mesh previously underwent the same biocompatibility testing as the permanent implant and met the predefined acceptance criteria.It is therefore a low risk material for biocompatibility.No non-conformances or complaints related to the mesh lot j02006 or sleeve lot j01006.Finished device passed all visual tip inspection, sleeve gap inspection, junction test, and mesh test, per lot history record.The subassembly sleeve width passed a sample inspection.Finally, the product was used prior to its expiration date of 02/27/2023.
 
Event Description
Dr.(b)(6) had an "assist" who was helping her today in her case (dr.(b)(6) wouldn't release if it was a resident or assistant).The person helping her cut the sheath right at the exit incision when dr.(b)(6) was tensioning her sling.The sheath was not held on to and was "sucked" back into the incision.The sheath was not in the patient and could not be retrieved.Dr.(b)(6) removed the placed sling in hopes the sheath would come out, but it didn't.Then surgeon tried to place another sling, by passing the trocar in hopes it would push the sheath out, it didn't.The new sling was taken out of the patient, and dr.(b)(6) called sales rep to discuss this issue.Sales rep mentioned to surgeon the "plastic sheath" would need to be removed from the patient's body.Later surgeon was able to dissect vaginally and pull out the sheath.Per dr.(b)(6), the complaint sling was disposed of after surgery and not available for return.No serious injury to the patient as a result of the procedure.No patient contraindications that could have caused or contributed to the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESARA
Type of Device
PELVIC MESH SLING
Manufacturer (Section D)
CALDERA MEDICAL INC.
5171 clareton drive
agoura hills CA 91301
Manufacturer (Section G)
CALDERA MEDICAL INC.
5171 clareton drive
agoura hills CA 91301
Manufacturer Contact
ieva sirokoviene
5171 clareton drive
agoura hills, CA 91301
8184837641
MDR Report Key8723448
MDR Text Key149231215
Report Number3003990090-2018-01454
Device Sequence Number1
Product Code OTN
UDI-Device Identifier00890594000049
UDI-Public00890594000049
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/27/2023
Device Catalogue NumberCAL-DS01
Device Lot NumberJ02006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2019
Initial Date FDA Received06/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-