Pinnacle claim submission has no allegation reported.After review of medical records, patient experienced discomfort and was revised to address pain and suspected altr reaction.Operative findings include thin greyish opaque fluid typical of an altr with mom bearings.This fluid was sent for cell count an resulted to 2450 white blood cells with 16% neutrophils.There was some black staining on the trunnion and a thin pseudocyst capsule.Blood metal ions are below 7.Doi: (b)(6) 2008 - dor: (b)(6) 2017, (right hip).
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|