Catalog Number 209999 |
Device Problems
Output Problem (3005); Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/10/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
Case number: (b)(4), mps (b)(6) reported no power to burr, dr requesting service visit.Case type: pka.Update: "<30 surgical delay".
|
|
Manufacturer Narrative
|
Reported event: mps (b)(6) reported no power to burr, dr requesting service visit.Device evaluation and results: per(b)(4): pka end effector magnet is weak and not connecting with burr motor.System investigation completed successfully as per service manual.All system checks and tests passed.Product history review: a review of device history records shows that on 11/03/2017 1 device was inspected and 1 device was placed on: qt 17-10-0102.A review of the data revealed that the non-conformances are not related to the failure alleged in this compliant.Complaint history review : a review of complaints in catsweb and trackwise related to p/n 209999, serial number (b)(6) shows no additional complaints related to the failure in this investigation.Conclusions: system ready for clinical use.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|
|
Event Description
|
Case number: (b)(4), mps (b)(6) reported no power to burr, dr requesting service visit.Case type: pka.Update: "<30 surgical delay".
|
|