MAUDE Adverse Event Report: SPACELABS HEALTHCARE INC. SPACELABS XHIBIT CENTRAL STATION; XHIBIT CENTRAL MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE INC. SPACELABS XHIBIT CENTRAL STATION; XHIBIT CENTRAL MONITOR Back to Search Results
Model Number 96102
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2019
Event Type  malfunction  
Manufacturer Narrative
Spacelabs has investigated this event and found that the reported event was likely to be caused by a faulty power supply.Spacelabs field service engineer has installed a loaner device for customer use and the original device will be sent in for repair.This report is complete and this particular issue is considered to be closed.
 
Event Description
Spacelabs received a report that the customer has experienced their telemetry central station powered down on its own on several occasions.On june 16th, monitoring was reported to be interrupted for 20 minutes.No injury was reported in association with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS XHIBIT CENTRAL STATION
Type of Device
XHIBIT CENTRAL MONITOR
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
beth benner rn
35301 se center st.
snoqualmie, WA 98065
4253635568
MDR Report Key8724796
MDR Text Key148972269
Report Number3010157426-2019-00023
Device Sequence Number1
Product Code MHX
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/01/2005,06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 06/05/2019
Initial Date FDA Received06/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-