• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. PATIENT CABLE; CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. PATIENT CABLE; CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR) Back to Search Results
Model Number 5487
Device Problems Failure to Capture (1081); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that no pacing occurred when the lead was plugged into the programmer cable.The lead also got stuck in the cable when it was attempted to be removed.The cable was replaced and there were no further issues.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: manufacturer's analysis was unable to confirm the customer comment that no pacing occurred when the lead was plugged into the programmer cable.The cable passed all functional testing.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The cable was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT CABLE
Type of Device
CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8730361
MDR Text Key149072330
Report Number2182208-2019-01130
Device Sequence Number1
Product Code DSA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960446
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5487
Device Catalogue Number5487
Device Lot Number686770001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/18/2019
Initial Date FDA Received06/25/2019
Supplement Dates Manufacturer Received07/03/2019
Supplement Dates FDA Received07/10/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age82 YR
-
-