Model Number M00510890
Device Problem Separation Failure (2547)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 05/27/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2019.According to the complainant, during the procedure, an alliance handle was used in conjunction with the trapezoid basket to crush a stone.However, the basket failed to crush the stone.Furthermore, the tip of the basket failed to detach to release the stone.The basket with the stone was finally removed by forcibly pulling the basket but caused bleeding.The bleeding was stopped by a metal stent, and the procedure was completed at this time.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
,
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
,
marlborough, MA 01752
5086834015
MDR Report Key8730885
MDR Text Key149078424
Report Number3005099803-2019-03252
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/05/2020
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0023445083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/28/2019
Initial Date FDA Received06/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-