Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/03/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device not returned for evaluation.
|
|
Event Description
|
It was reported that during a surgery, the impact plate fractured while impaction.No adverse events have been reported as a result of the malfunction.Additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.An offset stem driver was returned for evaluation.As returned, the strike plate has fractured from the main shaft.The material hardness of the shaft is conforming to print specifications.The device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information. reported event was confirmed by review of photographs provided.No product was returned, visual and dimensional evaluations could not be performed.However, photographs of the returned product were provided and it was identified the impactor plate had weld fracture from the handle.No other evaluation can be performed.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|