• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNIVERSAL INSERTER/EXTRACTOR; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNIVERSAL INSERTER/EXTRACTOR; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/10/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during an initial hip arthroplasty, the femoral impactors handle fractured.No adverse events have been reported as a result of the malfunction.Additional information on the reported event is unavailable.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Reported device was returned and evaluated.Upon visual inspection the device the black plastic had fractured.There is impact damage on the strike area along with both sides of the neck.The outside of the plastic sleeve also shows impact marks device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL INSERTER/EXTRACTOR
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8741530
MDR Text Key149517537
Report Number0001825034-2019-02755
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-473601
Device Lot NumberZB161201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/11/2019
Initial Date FDA Received06/27/2019
Supplement Dates Manufacturer Received10/01/2019
Supplement Dates FDA Received10/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-