Model Number 010315-
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/06/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during an abdominal aortic aneurysm procedure, while stapling the aorta, the stapler was unable to be released while clamped close.Device forced open with large amount of force.The staple line was intact and able to complete case.No patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted the instrument was received fully applied.The jaw was partially open.The instrument was melted and damaged from a sterilization process.The damage precludes functional evaluation.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the observed condition may occur when the instrument is autoclaved and/or a sterilization process.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM MULTIFIRE TA
Type of Device
APPARATUS, SUTURING, STOMACH AND INTESTINAL
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key8741833
MDR Text Key149493120
Report Number1219930-2019-03767
Device Sequence Number1
Product Code FHM
UDI-Device Identifier10884521045965
UDI-Public10884521045965
Combination Product (y/n)N
PMA/PMN Number
K801589
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number010315-
Device Catalogue Number010315-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Initial Date Manufacturer Received 06/07/2019
Initial Date FDA Received06/27/2019
Supplement Dates Manufacturer Received08/12/2019
Supplement Dates FDA Received09/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-