• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY PARADYM; DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY PARADYM; DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER Back to Search Results
Model Number PARADYM DR 8550
Device Problems Over-Sensing (1438); Low impedance (2285)
Patient Problem No Information (3190)
Event Date 06/06/2019
Event Type  Injury  
Event Description
Reportedly, during the follow-up on (b)(6) 2019, oversensing was observed on the ventricular channel in several episodes.The ventricular lead impedance curve was stable despite a slight decrease, however the coils continuity trends showed some values below 200 ohms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADYM
Type of Device
DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
SORIN GROUP ITALIA SRL VIA CRESCENTINO 13040 SALUGGIA ITALY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key8743318
MDR Text Key149488279
Report Number1000165971-2019-00380
Device Sequence Number1
Product Code MRM
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/15/2014
Device Model NumberPARADYM DR 8550
Device Catalogue NumberPARADYM DR 8550
Device Lot Number2687
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/06/2019
Event Location Hospital
Initial Date Manufacturer Received 06/06/2019
Initial Date FDA Received06/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-