• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND ARCHITECT TOTAL PSA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ARCHITECT TOTAL PSA Back to Search Results
Catalog Number 07K70-25
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.Note: this report is being filed on an international product, architect total psa ln 7k70 that has a similar product distributed in the us, list number 6c06.
 
Event Description
The customer observed falsely elevated total psa results for one patient on the architect i2000sr analyzer.The following data was provided for a post-op patient being monitored for prostate cancer: initial 0.47 ng/ml, repeat 0.Confirmed as 0.003 ng/ml on another method.There was no impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and instrument service.No adverse trend was identified for the customer's issue.A review of customer field data was performed which shows no atypical performance with reagent lot 03081fn00.A review using search of the corrective and preventive actions system and manufacturing documentation did not identify issues associated with the customer's observation.Return not available.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TOTAL PSA
Type of Device
TOTAL PSA
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
MDR Report Key8744191
MDR Text Key149992637
Report Number3008344661-2019-00079
Device Sequence Number1
Product Code LTJ
Combination Product (y/n)N
PMA/PMN Number
P910007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/26/2020
Device Catalogue Number07K70-25
Device Lot Number03081FN00
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/05/2019
Initial Date FDA Received06/28/2019
Supplement Dates Manufacturer Received09/09/2019
Supplement Dates FDA Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARCHITECT I2000SR ANALYZER; ARCHITECT I2000SR ANALYZER; LN 03M74-01 SN (B)(4) ; LN 03M74-01 SN (B)(4)
-
-