• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TAPERLOC POR RED/LAT 17.5X155; PROSTHESIS HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TAPERLOC POR RED/LAT 17.5X155; PROSTHESIS HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Material Erosion (1214)
Patient Problem Pain (1994)
Event Date 03/28/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical device: m2a-magnum pf cup 64odx58id, # item us157864, lot: 792860; m2a-magnum mod hd sz 58mm, # item 157458, lot: 832880; m2a-magnum 52-60mm tpr insrt-6, # item 139264, lot: 348620.Reported event was confirmed by review of op notes provided.Revision op notes dated were reviewed and identified patient was revised due to pain.During the revision, it was identified femoral head could not be disassociated from the stem with noted trunnionnosis.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient's right hip was revised approximately 7 years post implantation due to pain, discomfort, soreness, dysfunction, loss of range of motion and elevated metal ions.The surgeon found irritative joint fluid within the capsule when they opened the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERLOC POR RED/LAT 17.5X155
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8744445
MDR Text Key149511803
Report Number0001825034-2019-02823
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K030055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2009
Device Model NumberN/A
Device Catalogue Number13-103209
Device Lot Number479640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/04/2019
Initial Date FDA Received06/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-