Model Number 5436
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during an implant procedure, there was no signal from the leads.It was discovered that the connector of the patient cables were broken.It was recommended that the cables be returned for analysis, but its current status is unknown.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: manufacturer's analysis confirmed the customer comment that the patient cable had no signal.Due to damaged pins in the connector, the cable failed continuity testing through a standard connection.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The cables were returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANALYZER PATIENT CABLE
Type of Device
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8744698
MDR Text Key149523626
Report Number2182208-2019-01169
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5436
Device Catalogue Number5436
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2019
Initial Date FDA Received06/28/2019
Supplement Dates Manufacturer Received08/07/2019
Supplement Dates FDA Received08/14/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2090 PROGRAMMER, 229047 ANALYZER
-
-