• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ENCORE SHOULDER HUMERAL HEAD; HEAD, HUMERAL, NEUTRAL, 42-20

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ENCORE SHOULDER HUMERAL HEAD; HEAD, HUMERAL, NEUTRAL, 42-20 Back to Search Results
Model Number 520-42-020
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Pain (1994)
Event Date 05/29/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was due to pain.The previous surgery and the revision detailed in this investigation occurred over 5 years and 2 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the main contributor component listed in the complaint.The device and its applicable concomitant device was within their respective expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on- going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.There are multiple factors that may contribute to the event that are outside the control of djo surgical are soft tissue impingement, degenerative bone disease, excessive weight, patient's age, unbalanced joint or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - due to pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE SHOULDER HUMERAL HEAD
Type of Device
HEAD, HUMERAL, NEUTRAL, 42-20
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key8745948
MDR Text Key149565517
Report Number1644408-2019-00584
Device Sequence Number1
Product Code KWS
UDI-Device Identifier00888912025010
UDI-Public(01)00888912025010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model Number520-42-020
Device Catalogue Number520-42-020
Device Lot Number911C1016
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/29/2019
Initial Date FDA Received06/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/29/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
510-99-000, LOT 639G1043
Patient Outcome(s) Required Intervention;
Patient Age81 YR
-
-