• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem Defective Alarm (1014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/06/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the receiver had intermittent audio output.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The receiver was evaluated.An external visual inspection was performed and passed.Charge and boot testing was performed and passed.The receiver log was downloaded and reviewed, finding no errors related to the complaint.Functional testing was performed and passed all relevant testing.Audio\vibration test with global receiver communication tool test was performed and passed.Manual try-it audio\vibration testing was performed and passed.The receiver was opened and an internal visual inspection was performed and passed.Speaker audio and vibrator intermittent tests were performed and passed.Speaker and vibrator resistance were measured and were within specification.The allegation was not confirmed.A probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key8750707
MDR Text Key149854982
Report Number3004753838-2019-051211
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000153
UDI-Public10386270000153
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTK-DR-001
Device Lot Number5249466
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/09/2019
Initial Date FDA Received07/01/2019
Supplement Dates Manufacturer Received07/25/2019
Supplement Dates FDA Received08/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
Patient Weight69
-
-