• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Device Problem Device Alarm System (1012)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 01jul19.The manufacturer's technical services (ts) confirmed the reported error codes issue.They advised customer to try a known good battery and if that does not correct the problem to replace the ui board.Advised customer of the software needed and the need for an xp computer.Emailed customer link to incenter and provided eol letter.Advised customer that bench repair is available.Follow up attempts have been made with no response from the customer.Review of the service history indicates there have been no subsequent calls from the customer regarding this issue.If new information is received the case will be reopened and updated.
 
Event Description
The customer reported power supply issue error codes 3.3 v voltage rail failed & post 3.3 mon fail.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8751386
MDR Text Key149822037
Report Number2031642-2019-04187
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/19/2018
Initial Date FDA Received07/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-