Model Number 60-02-15
Device Problem Partial Blockage (1065)
Patient Problem No Patient Involvement (2645)
Event Date 06/03/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The affected part has been requested to return to livanova (b)(4) for a detailed investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that the shaft angle encoder of a centrifugal pump system with tubing clamp intermittently locked and did not turn during demonstration.There was no patient involvement.
 
Manufacturer Narrative
The affected parts were returned to livanova deutschland for a detailed investigation.The investigator could reproduce the reported issue.The problem could be traced to mechanical issues.Due to obsolescence of the part a new version of the encoder is used for the manufacturing of the new devices.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8752726
MDR Text Key149810758
Report Number9611109-2019-00509
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/03/2019
Initial Date FDA Received07/02/2019
Supplement Dates Manufacturer Received07/08/2019
Supplement Dates FDA Received08/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-