This event is recorded by zimmer biomet under (b)(4).On (b)(6) 2019, it was reported from (b)(6) hospital that unit was leaking fluid from both right and left manifold housings.The zimmer biomet authorized service technician from (b)(4) arrived at the site on 01 july 2019, evaluated the unit and confirmed the reported event.After that he replaced manifold housings and confirmed that the unit was functioning as intended.The service technician then returned the unit back to service without any further concerns.The unit was tested, inspected, and repaired.While the service technician confirmed the reported event and the device was noted to be functioning as intended after the cl 500 housing were replaced on both the cylinders of cart, it is unknown with the information as why there was a leak from the cl500 housing.Therefore, based on the information provided, a specific root cause of the reported event cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Product evaluated by external contractor.
|