Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 03jul2019.The manufacturer's field service engineer (fse) performed remote troubleshooting and advised the customer that the user interface (ui) may be faulty.The fse recommended bench repair for the unit.The customer was advised to call back should bench repair be desired.The customer did not respond to follow up message and emails regarding the status of the repairs.Review of the service history indicates there have been no subsequent calls from the customer regarding this issue.If new information is received the case will be reopened and updated.
 
Event Description
The customer reported that the unit will not power on.The device was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or user.
 
Manufacturer Narrative
Is device single use?: reuse.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8760109
MDR Text Key150149878
Report Number2031642-2019-04219
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/10/2018
Initial Date FDA Received07/03/2019
Supplement Dates Manufacturer Received01/10/2018
Supplement Dates FDA Received07/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-