We were notified of the following event: "information received by email on june 6, 2019 from area rep c/o customer (b)(6) (email ccd) (b)(6) stating "the following account has a scale calibration issue that needs to be repaired immediately.We were in a hysteroscopy case yesterday (b)(6) 2019.During the case, the deficit fluctuated from a positive number all the way up to 210 down to a negative 180.Intermittently during the case.A test was run on the unit and it ran negative in the end as well.Please have a replacement unit sent to this facility asap as they are down to one usable unit.(b)(6).(b)(4)."".
|