Catalog Number 72200568S
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that during surgery the device powered off; however, the device's fan did not stop.It is unknown whether there was a backup device available.No delay or patient injury was reported.
 
Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of power failure could not be confirmed.Product passed all functional testing.All functions perform as expected and no loss of power occurred during testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key8773639
MDR Text Key150545444
Report Number1643264-2019-00455
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568S
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/18/2019
Initial Date FDA Received07/09/2019
Supplement Dates Manufacturer Received08/09/2019
Supplement Dates FDA Received08/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-