• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC ESPRIT VENTILATOR Back to Search Results
Model Number V1000
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 10july2019.A follow-up report will be submitted once the investigation has been complete.
 
Event Description
The customer reported display issue.Display is missing format and intelligence and a splash of red/orange is present.There was no patient involvement.
 
Manufacturer Narrative
Date rec¿d by mfr: 29oct2019.Date of report: 30oct2019.The manufacturer¿s technical services (ts) confirmed the reported issue.The customer has received the replacement vga pcb but has not installed it at this time.The customer had declined to repair the unit at this time.If further information is obtained, this complaint will be reopened.The determination could not be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT VENTILATOR
Type of Device
VENTILATOR
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8776098
MDR Text Key150723550
Report Number2031642-2019-04389
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K981072
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV1000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/25/2019
Initial Date FDA Received07/10/2019
Supplement Dates Manufacturer Received06/25/2019
Supplement Dates FDA Received10/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-