• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS Back to Search Results
Catalog Number 966180
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Bolts providing secondary fixation of femoral impactor head loose to the point of complete separation of components.Threads not damaged and able to reassemble during post-operative washup.Complaint not entered immediately as thought to be simply disassembled rather than broken; request for product complaint from product specialist.Was surgery delayed due to the reported event? --> no.Was procedure successfully completed? --> yes.Were fragments generated? --> no.Patient status/ outcome / consequences --> no.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).The device was reviewed by depuy engineering melbourne in a- 2895204 which states my assessment is that this complaint is due to wear and tear.Root cause attributed to the device being worn from normal use and servicing.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8776102
MDR Text Key150573028
Report Number1818910-2019-97947
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295247203
UDI-Public10603295247203
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number966180
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/24/2019
Initial Date FDA Received07/10/2019
Supplement Dates Manufacturer Received09/02/2019
Supplement Dates FDA Received09/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-