• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; INTRODUCER, CATHETER Back to Search Results
Model Number 406849
Device Problem Material Perforation (2205)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/29/2019
Event Type  malfunction  
Manufacturer Narrative
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.The device was returned due to brk needle insertion difficulty.The sheath had been perforated proximal to the distal tip.A needle/stylet assembly from current inventory was inserted and advanced through the dilator; however, resistance was noted at the aforementioned perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the insertion difficulty and perforation remains unknown.
 
Event Description
This report is to advise of an event observed during analysis confirming a punctured introducer sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8777921
MDR Text Key150820524
Report Number3005334138-2019-00362
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734203388
UDI-Public05414734203388
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K061015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model Number406849
Device Lot Number6675748
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2019
Initial Date FDA Received07/10/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-