• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIMACORPORATE SPA SMR - EXPANSION EXTRACTOR; SMR - EXPANSION EXTRACTOR (KWT, HSD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIMACORPORATE SPA SMR - EXPANSION EXTRACTOR; SMR - EXPANSION EXTRACTOR (KWT, HSD) Back to Search Results
Model Number 9013.52.165
Device Problems Output Problem (3005); Physical Resistance/Sticking (4012); Difficult or Delayed Separation (4044)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/04/2019
Event Type  malfunction  
Manufacturer Narrative
By checking the manufacturing chart of the involved lot numbers, no anomaly was found.This is the first and only complaint received on these lot numbers and codes.We will submit a final mdr once the investigation will be completed.
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During surgery, extraction instruments for reverse body malfunctioned and failed to dissociate reverse humeral body from the stem.The event prolonged the surgery of 15 minutes.The instruments involved are the following: 9013.50.175, smr - univer.Stem f.Extractor, lot #14aa136.9013.52.165, smr - expansion extractor, lot #14aa416.According to the info reported, it is not clear which instrument did not work correctly.Surgeon completed the surgery using the instruments from the anatomical smr tray.Event occurred in (b)(6).
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During surgery, extraction instruments for reverse body malfunctioned and failed to dissociate reverse humeral body from the stem.The event prolonged the surgery of 15 minutes.The following instruments were involved: 9013.50.175, smr - univer.Stem extractor, lot #:14aa136; 9013.52.165 smr - expansion extractor lot #:14aa416.According to the info reported, it is not clear which instrument did not work correctly.Surgeon completed the surgery using the instruments from the anatomical smr tray.Event occurred in australia.
 
Manufacturer Narrative
By checking the dhrs of the involved lots, no anomaly was found.On a total of (b)(4)expansion extractors manufactured with lot #14aa416.On a total of (b)(4) stem extractors manufactured with lot #14aa136.We requested the instruments to be returned for analysis, but they were not sent to us.At this stage, we are not able to investigate deeply the root cause of the event.On the basis of the check of the manufacturing chart, we can judge this event as not product related.Pms data: (b)(4).Please note that occurence rate is highly overestimated since it is not possible to take into account the number of usage of these instruments (re-usable components).No corrective actions planned for this specific case.Lima corporate will keep the market monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR - EXPANSION EXTRACTOR
Type of Device
SMR - EXPANSION EXTRACTOR (KWT, HSD)
Manufacturer (Section D)
LIMACORPORATE SPA
via nazionale 52
villanova di san daniele, 33038
IT  33038
MDR Report Key8788219
MDR Text Key200417224
Report Number3008021110-2019-00079
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K100858
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number9013.52.165
Device Lot Number14AA416
Initial Date Manufacturer Received 07/05/2019
Initial Date FDA Received07/15/2019
Supplement Dates Manufacturer Received07/05/2019
Supplement Dates FDA Received06/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-