Japan warehouse complaint of "black particulate" was confirmed.As received, an unknown loose black particulate was found already isolated from the cannula.As received, the device appeared to have incomplete printing on the wired reinforced section of the cannula.Upon examination, the non-wired section of the cannula appeared to have unknown contaminates.Device was sent to chemistry lab for ft-ir analysis.No other visual damage, contamination, or other abnormalities were found to the device.Engineering task has been opened and assigned for further investigation.A supplemental report will be submitted.
|
Device was sent to chemistry lab for ft-ir analysis.The particulate was removed from the sample slide and scanned by diamond atr.Particulate was broken into two pieces, measuring at 1674 m by 640 m for an area measurement of 1,071,360 m² and 1446 m by 678 m for an area measurement of 980,388 m².The ft-ir spectrum showed similar absorption characteristics to acrylic polymer like material.The complaint is confirmed.Per the product evaluation and chemistry evaluation, the complaint of "black particulate" was confirmed.Per the engineering evaluation, a manufacturing defect was confirmed.A supplier, design, ifu, and labeling defect were not confirmed.The trend is in control and the fmea line item is appropriate.Pra was initiated and the root cause will be investigated in capa.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|