• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/27/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4).In the confirmation at the service station of osh, the reported phenomenon was duplicated.In addition, the following were confirmed; failure of the remote switch.Scope communication error e30.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the image of the subject device became black and white stripes at pre-use inspection at the user facility.The user facility canceled the procedure.There was no patient injury associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to additional information.Olympus medical systems corp.(omsc) reviewed the manufacture history (dhr) of the device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.The exact cause could not be determined, but it was surmised that the phenomenon occurred due to ccd damage from the result of confirmation by osh.Ccd damage may have occurred due to excessive bending and tensile stress applied to the distal end and the bending section where the ccd was built.The instruction manual of the subject device already instructs; follow the warnings given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.Do not bend, hit, pull, or twist the insertion section, bending section, control section, universal cord, video cable, video connector, and light guide connector.The endoscope may be damaged, and water leakage and/or breakage of internal parts like the image sensor cable can result.If any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2, ¿troubleshooting guide¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8796819
MDR Text Key151315911
Report Number8010047-2019-02552
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/01/2019
Initial Date FDA Received07/17/2019
Supplement Dates Manufacturer Received12/03/2019
Supplement Dates FDA Received12/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-