MAUDE Adverse Event Report: RADIOMETER MEDICAL APS ABL90 FLEX PLUS ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RADIOMETER MEDICAL APS ABL90 FLEX PLUS ANALYZER Back to Search Results
Model Number 393-092
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/03/2019
Event Type  Injury  
Manufacturer Narrative
The hospital staff needed to perform a medical intervention and terminate the inappropriate ca2+ treatment.Due to the ca2+ treatment, the patient developed moderate hypercalcemia.However, no further complications for the patient was associated with this event.Root cause: data logs have been reviewed and from the investigation it is suggested that a clot was introduced to the instrument at a time in between the (b)(6) 2019 01:02 and (b)(6) 2019 05:01 thus giving a lower than expected measurement on ca2+ on one or more of the measurements performed after the time of introduction.This is considered a preanalytic error.
 
Event Description
According to the complaint, a customer experienced several low measurements on ionized ca2+ (ica) on samples from a (b)(6) year old female patient using an abl90 flex analyzer ((b)(4)) during citrate dialysis.Before citrate dialysis the following result for ica was measured: (b)(6) 2019 19:27h : 1.13 mmol/l (ica patient).Hereafter the citrate dialysis was initiated and the following results were measured for ica: (b)(6) 2019 23:30h : 0.40 mmol/l (post filter); (b)(6) 2019 23:31h : 1.03mmol/l (ica patient).Based on the results, the doctor initiated ca2+ treatment (150%) to the patient and the following results were measured for ica: (b)(6) 2019 00:38h : 0.43 mmol/l (post filter); (b)(6) 2019 00:40h : 1.00 mmol/l (ica patient); (b)(6) 2019 02:19h : 1.01 mmol/l (ica patient); (b)(6) 2019 02:20h : 0.41 mmol/l (ica patient); (b)(6) 2019 03:37h : 0.78 mmol/l (ica patient); (b)(6) 2019 03:38h : 1070 sensor response error.Unstable signal from sensor.Repeat measurement (post filter); (b)(6) 2019 03:39h : 0.4 mmol/l (post filter).The doctor increased the ca2+ treatment (180%) and the following results were measured for ica: (b)(6) 2019 04:46h : 0.88 mmol/l (ica patient); (b)(6) 2019 04:47h : 0.52 mmol/l (ica patient); (b)(6) 2019 04:48h : 0.51 mmol/l (post filter); (b)(6) 2019 05:50h : 1070 sensor response error.Unstable signal from sensor.Repeat measurement (ica patient).The doctor assessed that the treatment was abnormal and performed a medical intervention to terminate the treatment.Due to the ca2+ treatment, the patient developed moderate hypercalcemia.No further complications for the patient was associated with this event.Subsequent to the citrate dialysis two comparison measurements were performed in the laboratory that gave the following results for ica: (b)(6) 2019 06:13h : 1.60 mmol/l (ica patient) abl825 flex (abl2 labo); (b)(6) 2019 06:23h : 1.62 mmol/l (ica patient) abl90 flex (abl1 labo).
 
Manufacturer Narrative
It was identified that an incorrect model number was stated in initial mdr.The brand name (field d1), common name (field d2), including product code, model #, catalogue # (field d4) and pma/510(k) (field g5) have all been updated accordingly in this follow-up #1 report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL90 FLEX PLUS ANALYZER
Type of Device
ABL90 FLEX PLUS ANALYZER
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA  2700
MDR Report Key8797724
MDR Text Key151286737
Report Number3002807968-2019-00033
Device Sequence Number1
Product Code MQM
UDI-Device Identifier05700693930923
UDI-Public(01)05700693930923
Combination Product (y/n)N
PMA/PMN Number
K160153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number393-092
Device Catalogue Number393-092
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/04/2019
Initial Date FDA Received07/17/2019
Supplement Dates Manufacturer Received09/20/2019
Supplement Dates FDA Received09/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
-
-