ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR
|
Back to Search Results |
|
Model Number AU00T0 |
Device Problems
Entrapment of Device (1212); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
|
|
Event Description
|
A health professional reported that during an intraocular lens (iol) implant surgery, the lens did not come out of the injector.The device was exchanged to complete the case.There was no patient impact.
|
|
Manufacturer Narrative
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A non-qualified viscoelastic was indicated.The root cause may be related to a failure to follow the dfu.A non-qualified viscoelastic was used in the device.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.As stated in the company aspheric intraocular lens with company preloaded delivery system dfu, only qualified company viscoelastics must be used in conjunction with the device.All other viscoelastics are not qualified for use.Lens delivery performance may be negatively affected when using other non-qualified viscoelastics leading to lens delivery issues and /or damage.The use of non-qualified viscoelastics is considered a failure to follow the dfu for the company aspheric intraocular lens with company preloaded delivery system and is not recommended under any circumstance.The manufacturer internal reference number is: (b)(4).
|
|