• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE SECTOR II CUP 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE SECTOR II CUP 48MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121722048
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Following the implantation of a 48 mm pinnacle sector acetabular cup, and during the insertion the apex hole eliminator it did not engage in the cup and went through the apical hole.The patient was treated with total hip replacement due to severe osteoarthritis of the right hip on (b)(6) 2019.The insertion of the acetabular cup pinnacle 48mm was uneventful.A failed engagement between the apex hole eliminator and the threaded part of the apical hole was noticed.The complete advancement of the apex hole eliminator through the hole resulted finally in its entrapment behind the posterior cup surface.At the immediate post-operative period, no complications were recorded.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found one other reported incident against the provided product/lot combination, (b)(4).Both complaints were from the same hospital and the same surgeon.Corrective action was not indicated.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE SECTOR II CUP 48MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8808272
MDR Text Key151619112
Report Number1818910-2019-98825
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295009801
UDI-Public10603295009801
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number121722048
Device Lot Number9159557
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/01/2019
Initial Date FDA Received07/19/2019
Supplement Dates Manufacturer Received09/23/2019
Supplement Dates FDA Received09/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-