TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number 1000096 |
Device Problems
Pumping Stopped (1503); Battery Problem (2885); Material Deformation (2976)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 06/29/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received.
|
|
Event Description
|
It was reported that the pump battery was observed to be depleting rapidly.Subsequently the pump shut down.Additionally, the backside of the pump where the battery is located was observed to have "expanded" and the customer was unable to turn the pump on again.The customer's blood glucose (bg) was 600 mg/dl.The customer declined to provide information regarding how the elevated bg was addressed or the form of alternate insulin therapy that the customer reverted to.
|
|
Manufacturer Narrative
|
The investigation has been completed.Based on the analysis, the alleged battery issue was verified; however, the 'expanded battery' was not verified.Additionally, a different issue was identified (shattered touchscreen).
|
|