• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number 1000096
Device Problems Pumping Stopped (1503); Battery Problem (2885); Material Deformation (2976)
Patient Problem Hyperglycemia (1905)
Event Date 06/29/2019
Event Type  Injury  
Manufacturer Narrative
The device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received.
 
Event Description
It was reported that the pump battery was observed to be depleting rapidly.Subsequently the pump shut down.Additionally, the backside of the pump where the battery is located was observed to have "expanded" and the customer was unable to turn the pump on again.The customer's blood glucose (bg) was 600 mg/dl.The customer declined to provide information regarding how the elevated bg was addressed or the form of alternate insulin therapy that the customer reverted to.
 
Manufacturer Narrative
The investigation has been completed.Based on the analysis, the alleged battery issue was verified; however, the 'expanded battery' was not verified.Additionally, a different issue was identified (shattered touchscreen).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
julia kensick
11075 roselle st.
san diego, CA 92121
8583666900
MDR Report Key8812313
MDR Text Key151747590
Report Number3013756811-2019-42976
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00853052007981
UDI-Public(01)00853052007981
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000096
Device Catalogue Number1003808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/29/2019
Initial Date FDA Received07/22/2019
Supplement Dates Manufacturer Received09/10/2019
Supplement Dates FDA Received09/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age42 YR
-
-